Comparison of dedicated breast positron emission tomography and whole-body positron emission tomography/computed
Yoko Satoh1,2, Utaroh Motosugi3, Masamichi Imai4
1Yamanashi PET Imaging Clinic, Shimokato 3046-2, Chuo, Yamanashi, 409-3821, Japan. ysatoh@ypic.jp.
Objective:
High-resolution dedicated breast positron emission tomography (dbPET) can visualize breast cancer more clearly than whole-body PET/computed tomography (CT). In Japan, the combined use of dbPET and whole-body PET/CT is necessary in indications for health insurance. Although several clinical studies have compared both devices, a physical evaluation by the phantom test has not been reported. The aim of this study was to compare the ability of ring-shaped dbPET and whole-body PET/CT using a common phantom with reference to the Japanese guideline for the oncology 18F-fluorodeoxyglucose (FDG)-PET/CT data acquisition protocol.
Methods:
A cylindrical breast phantom with four spheres of different diameters (16, 10, 7.5, and 5 mm) filled an FDG solution at sphere-to-background radioactivity ratios (SBRs) of 2:1, 4:1, and 8:1 was prepared. Images were then acquired by whole-body PET/CT and subsequently by dbPET. The reconstructed images were visually evaluated and the coefficient of variation and uniformity of the background (CVbackground and SDΔSUVmean), percentages of contrast and background variability (%QH,5mm and %N5mm), and their ratio (%QH,5mm/N5mm), and relative recovery coefficient were compared with the standards defined in the protocol for whole-body PET/CT.
Results:
The parameters were calculated at an SBR of 8:1, which was the only SBR in which a 5-mm sphere was visible on both devices. The standards were defined as < 10% for CVbackground, ≤ 0.025 for SDΔSUVmean, < 5.6% for %N5mm, > 2.8 for %QH,5mm/N5mm, and > 0.38 for the relative recovery coefficient of the smallest sphere (10 mm in diameter) in the protocol for whole-body PET/CT (the %QH,5mm was not determined for that protocol); the respective values were 6.14%, 0.024, 4.55%, 3.66, and 0.33 for dbPET and 2.21%, 0.021, 3.11%, 1.72, and 0.18 for PET/CT. The QH,5mm was 16.67% for dbPET and 5.34% for PET/CT. The human images also showed higher lesion-to-background contrast on dbPET than on PET/CT despite the noisier background observed with dbPET.
Conclusion:
The common phantom study showed that the background was noisier and that the contrast was much higher in the dbPET image than in the PET/CT image. The acquisition protocol and standards for dbPET will need to be different from those used for whole-body PET/CT.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...