X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Occupational Doses to Medical Staff Performing or Assisting with Fluoroscopically Guided Interventional Procedures
David Borrego1, Cari M Kitahara1, Stephen Balter1
1From the Radiation Epidemiology Branch, Division of Cancer Epidemiology and Genetics, National Cancer Institute, National Institutes of Health, 9609 Medical Center Dr, Bethesda, MD 20892-9778 (D.B., C.M.K.); Departments of Radiology and Medicine, Columbia University Medical Center, New York, NY (S.B.); and Independent consultant, Weddington, NC (C.Y.).
Abstract:
Background Staff who perform fluoroscopically guided interventional (FGI) procedures are among the most highly radiation-exposed groups in medicine. However, there are limited data on monthly or annual doses (or dose trends over time) for these workers. Purpose To summarize occupational badge doses (lens dose equivalent and effective dose equivalent values) for medical staff performing or assisting with FGI procedures in 3 recent years after accounting for uninformative values and one- versus two-badge monitoring protocol. Materials and Methods Badge dose entries of medical workers believed to have performed or assisted with FGI procedures were retrospectively collected from the largest dosimetry provider in the United States for 49 991, 81 561, and 125 669 medical staff corresponding to years 2009, 2012, and 2015, respectively. Entries judged to be uninformative of occupational doses to FGI procedures staff were excluded. Monthly and annual occupational doses were described using summary statistics. Results After exclusions, 22.2% (153 033 of 687 912) of the two- and 32.9% (450 173 of 1 366 736) of the one-badge entries were judged to be informative. There were 335 225 and 916 563 of the two- and one-badge entries excluded, respectively, with minimal readings in the above-apron badge. Among the two-badge entries, 123 595 were incomplete and 76 059 had readings indicating incorrect wear of the badges. From 2009 to 2015 there was no change in lens dose equivalent values among workers who wore one badge (P = .96) or those who wore two badges (P = .23). Annual lens dose equivalents for workers wearing one badge (median, 6.9 mSv; interquartile range, 3.8213.8 mSv; n = 6218) were similar to those of staff wearing two badges (median, 7.1 mSv; interquartile range, 4.6-11.2 mSv; n = 1449) (P = .18), suggesting a similar radiation environment. Conclusion These workers are among the highest exposed to elevated levels of ionizing radiation, although their occupational doses are within U.S. regulatory limits. This is a population that requires consistent and accurate dose monitoring; however, failure to return one or both badges, reversal of badges, and improper badge placement are a major hindrance to this goal. © RSNA, 2019 Online supplemental material is available for this article. See also the editorial by Karellas in this issue.
More Related Videos
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography