Typical BBB morphology and implantation depth of 3830 electrode predict QRS correction by left bundle branch area
Zhixin Jiang1, Qin Chang1, Yucheng Wu1
1Department of Cardiology, The First Affiliated Hospital of Nanjing Medical University, Jiangsu Province Hospital, China.
Aim:
Strict criteria of typical left bundle branch block (LBBB) can help with the prediction for cardiac resynchronization therapy response. The aim of this study is to determine whether the use of strict criteria for both LBBB and right bundle branch block (RBBB) predicts successful QRS correction (≤130 ms) by left bundle branch area pacing (LBBAP).
Methods:
Consecutive patients with pacemaker indications according to the present guideline who also underwent LBBAP implantation were retrospectively assessed. Inclusion criteria were patients with BBB and the baseline QRSd > 130 ms. Baseline characteristics and pacing parameters were compared between typical and atypical BBB groups. Multivariate logistic regression was used to adjust for covariates that were found in univariate analyses for successful QRS correction by LBBAP.
Results:
Seventy-three patients were enrolled. Among them, 10 (13.6%) had atypical BBB (5 LBBB and 5 RBBB) and 63 (86.4%) had typical BBB (30 LBBB and 33 RBBB). The rate of successful QRS correction was higher in typical-BBB patients (52/63; 82.5%) than that in atypical-BBB patients (3/10; 30%), P < .001. Paced QRSd was obviously narrower in patients with typical BBB than that in patients with atypical-BBB (118 ± 14 vs 133 ± 14 ms, P = .003). In multivariate logistic regression, only typical BBB morphology and the implantation depth of 3830 pacing electrode in the ventricular septum were independent predictors for successful QRS correction.
Conclusion:
This study demonstrates that patients with typical-BBB morphology benefit more from LBBAP for QRS correction. Typical BBB morphology together with deep penetration of 3830 ventricular electrode in the interventricular septum predicts the success of QRS correction by LBBAP.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Bode Plots Construction
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation