Assessment of unintended volume loss of six closed system transfer devices
Stephen Kulju1, Bruce A McIntosh1, Helen Ja Fuller1
1VA National Center for Patient Safety, Ann Arbor, MI, USA.
Introduction:
Closed system transfer devices (CSTDs) are used to prepare and administer hazardous drugs. Previous studies have explored the vapor and fluid containment performance of CSTDs. A less obvious consideration is the effect of CSTD use on the intended dose for small volume administrations. We assessed six CSTDs to determine if they contribute to volume loss and delivery of less than the intended dose during simulated drug administration.
Methods:
Using an analytical balance, we obtained the mass of each CSTD at four points during simulated drug preparation and subcutaneous administration using sterile water. We used the masses to determine the average volume loss (V) for each CSTD.
Results:
Using ANOVA, we identified significant differences in volume loss (VL) among the mean VL (F(6,59) = 18.45, p = 6.19 × 10-12) for the six CSTDs and control (no CSTD). Four CSTDs had a VL that was statistically different than the control (p < 0.05); the VL for two CSTDs was not statistically different than the control (p > 0.05). Volume loss did not depend on administration volume.
Conclusion:
Volume loss performance varied among CSTDs. Volume loss may be clinically significant for small volumes but is not likely clinically significant for larger volumes. It is the authors' opinion that 3 mL represents a reasonable administration threshold below which volume loss should be considered clinically significant. Users should consider volume loss in context of the tasks, drugs, users, and environments where CSTDs will be used. The United States Pharmacopeia (USP) General Chapter <800>: Hazardous Drugs-Handling in Healthcare Settings recognizes the lack of CSTD performance standards. USP <800> recommends independent performance evaluation of CSTDs based on peer-reviewed studies. Our study adds to the comparative performance evaluations of CSTDs available on the market at the time of this review.
Related Concept Videos
Minor Losses in Pipes
Valves play a significant role in generating minor losses by obstructing or redirecting the fluid flow. When a valve is closed or partially closed, it restricts the flow...
Pipe Flowrate Measurement
The orifice meter is a simple,...
Pipe Flowrate Measurement: Problem Solving
Design Example: Designing a Residential Plumbing System
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Urodynamic Studies: Uroflowmetry