A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
A novel fiber-optic pressure sensor device for measuring variceal pressure
Xi Wang1, Huixian Li1, Jing Wang1
1Department of Gastroenterology, Anhui Geriatric Institute, The First Affiliated Hospital of Anhui Medical University, Hefei, Anhui 230022, P.R. China.
A new endoscopic fiber-optic pressure sensor accurately measures variceal pressure, predicting bleeding risk in cirrhosis patients. This convenient method offers a reliable alternative to invasive techniques for assessing variceal bleeding risk.
Area of Science:
- Gastroenterology
- Medical Devices
- Surgical Technology
Background:
- Variceal pressure is a key predictor of variceal bleeding.
- Current methods for measuring variceal pressure are often invasive or inconvenient.
- There is a need for a reliable and feasible method to assess variceal pressure.
Purpose of the Study:
- To develop and evaluate an endoscopic fiber-optic pressure sensor for measuring variceal pressure.
- To assess the accuracy, reliability, and feasibility of this novel sensor in a clinical setting.
- To determine if the sensor can effectively predict the risk of variceal bleeding.
Main Methods:
- A prospective study involving 80 cirrhosis patients with a history of variceal bleeding.
- An endoscopic fiber-optic pressure sensor was used to measure variceal pressure via the endoscope's biopsy channel.
- Variceal pressure measurements were compared with the portal pressure gradient (PPG) measured by transjugular intrahepatic stent-shunt (TIPS).
Main Results:
- The fiber-optic pressure sensor system was technically satisfactory in 78 out of 80 patients.
- A strong linear correlation was observed between variceal pressure measured by the sensor and PPG (r=0.940, P<0.001).
- The sensor demonstrated accuracy and feasibility in clinical measurements.
Conclusions:
- The endoscopic fiber-optic pressure sensor is an accurate and feasible tool for measuring variceal pressure.
- This technology effectively predicts the risk of variceal bleeding in cirrhosis patients.
- The sensor offers a promising, less invasive alternative for managing patients at risk of variceal hemorrhage.
More Related Videos
09:04Epidural Intracranial Pressure Measurement in Rats Using a Fiber-optic Pressure Transducer
Published on: April 25, 2012
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Pressure Gauges
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.