Design and fabrication of a new multi-loop saddle coil for 1.5 T MRI
1Department of Energy Engineering, Sharif University of Technology, Tehran 1458889694, Iran.
Abstract:
Radiofrequency coils provide high-resolution magnetic resonance (MR) imaging of human tissues. A small RF coil produces MR images with a higher resolution compared to the commercial volume MR coils from mass limited samples. Signal to noise ratio (SNR) plays a key role in the optimal design of receiver radiofrequency coils. In this work, we present a three-loop saddle coil suitable for MR imaging of digits of the human body. The geometry of the introduced coil is optimized to achieve the highest SNR. The coil performance is evaluated through comparing the measured SNR maps of the optimal coil derived from MR images of a saline phantom with the corresponding measured SNR maps of a commercial head coil in axial and sagittal slices. Results verify that the image SNR of the introduced coil is 3.4 times higher than that of the head coil and 2 times higher than that of the similar saddle coils represented in the literature recently. To validate the measured results, SNR maps of the introduced saddle and head coils were simulated and their SNR difference was compared with the corresponding measured data of the two coils. Results show that the simulated and measured data are in agreement with less than 11.8% error.
Related Concept Videos
Magnetic Field Of A Current Loop
Torque On A Current Loop In A Magnetic Field
Consider a rectangular current-carrying loop containing N turns of wire, placed in a uniform magnetic field. The net force on a current-carrying loop...
Magnetic Field Due to Two Straight Wires
Magnetic Resonance Imaging
Magnetic Field Due To A Thin Straight Wire
Magnetic Field of a Solenoid
Consider a solenoid with 100 turns wrapped around a cylinder of...