Reduction of the fluorine-18-labeled fluorodeoxyglucose dose for clinically dedicated breast positron emission
Yoko Satoh1,2, Tetsuro Sekine3, Yoshie Omiya4
1Yamanashi PET Imaging Clinic, Shimokato 3046-2, Chuo City, Yamanashi Prefecture, 409-3821, Japan. ysatoh@ypic.jp.
Purpose:
To determine the clinically acceptable level of reduction in the injected fluorine-18 (18F)-labeled fluorodeoxyglucose (18F-FDG) dose in dedicated breast positron emission tomography (dbPET).
Methods:
A breast phantom with four spheres exhibiting various diameters (5, 7.5, 10, and 16 mm), a background 18F-FDG radioactivity of 2.28 kBq/mL, and a sphere-to-background radioactivity ratio of 8:1 was used. True dose-reduced dbPET images were obtained by data acquisition for 20 min in list mode at multiple time points over 7 h of radioactive decay. Simulated dose-reduced images were generated by reconstruction with a portion of the list mode acquisition data. True and simulated dose-reduced images were visually and quantitatively compared. On the basis of the phantom study, dbPET images for 32 breasts of 28 women with abnormal uptake were generated after simulated reduction of the injected 18F-FDG doses; these images were compared with those acquired using current clinical doses.
Results:
There were no qualitative differences between true and simulated dose-reduced phantom images. The phantom study revealed that the minimal required dose was 12.5% for the detection of 5-mm spheres and 25% for precise semi-quantification of FDG in the spheres. The 7-min reconstruction with a 100% dose was defined as the reference for the clinical study. The image quality and lesion conspicuity were clinically acceptable for the 25% dose images. Lesion detectability on the 12.5% dose images was maintained despite image quality degradation.
Conclusions:
In summary, 25% of the standard 18F-FDG dose for dbPET can provide a clinically acceptable image quality, while 12.5% of the standard dose results in acceptable quality in terms of lesion detection when lesions are located at a sufficient distance from the edge of the dbPET detector.
More Related Videos
09:03Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET