Evaluation of sample stability for cellular kinetics and pharmacodynamic flow cytometry methods
Related Experiment Video
Updated: Jan 2, 2026

Measuring Cell Cycle Progression Kinetics with Metabolic Labeling and Flow Cytometry
Published on: May 22, 2012
Vellalore N Kakkanaiah1, Fan Pan1, Patrick Bennett1
1Biomarker Laboratories, PPD, Richmond, VA 23230, USA.
Abstract:
We evaluated the sample stability for a cellular kinetics and a pharmacodynamic flow cytometry methods. First, the blood collection tubes were compared for the enumeration of chimeric antigen receptor-T cells in human whole blood. Blood samples with chimeric antigen receptor-T cells were stable up to 3 days at room temperature in both conventional EDTA and Cyto-Chex® blood collection tubes (Streck Laboratories, NE, USA), but with better consistency in Cyto-Chex-BCT than conventional EDTA tubes. Second, sample storage temperatures were compared for the basophil activation test in human whole blood samples. The samples were stable up to 3 days for basophil activation test when stored at refrigerator temperature, but not stable when stored at room temperature. It is crucial during the development of method to evaluate all the variables which might impact sample integrity.
More Related Videos
09:57Characterizing Microbiome Dynamics – Flow Cytometry Based Workflows from Pure Cultures to Natural Communities
Published on: July 12, 2018
11:32Determination of the Relative Cell Surface and Total Expression of Recombinant Ion Channels Using Flow Cytometry
Published on: September 28, 2016