Using a TRAPS upstream transmission detector to verify multileaf collimator positions during dynamic radiotherapy
L Beck1, J J Velthuis2, S Fletcher3
1University of Bristol, H.H. Wills Physics Laboratory, Tyndall Avenue, BS8 1TL, Bristol, United Kingdom.
None:
With the advancement of high-precision radiotherapy and the increasing use of higher intensity beams, the risk to the patient increases should the radiotherapy machine malfunction. Hence more accurate treatment verification is required. In this paper we provide a solution for real-time monitoring of X-ray beams from radiotherapy linear accelerators using monolithic active pixel sensors. We show that leaf errors can be detected with high precision in static fields and IMRT step and shoot, and accurate leaf tracking is possible in Volumetric Modulated Arc Therapy. The prototype MAPS detector meets the criteria of 1% attenuation acceptable for clinical use.