Comparison of Three Point-of-Care Ultrasound Views and MRI Measurements for Optic Nerve Sheath Diameter: A
Ronak Raval1, Jay Shen2, Deon Lau1
1Department of Anesthesiology, Loma Linda University Medical Center, Loma Linda, CA, USA.
Introduction:
Point-of-care ultrasound of the optic nerve sheath diameter (ONSD) to diagnose increased intracranial pressure (ICP) is of great interest in various clinical scenarios. Yet, the lack of examination standardization has made clinical utility difficult. We compare three ultrasound ocular plane views (inferior, sagittal, and transverse), which are currently used in the literature to evaluate their consistency. Comparisons for each view to magnetic resonance imaging (MRI) measurements were also made.
Methods:
Fifty-one patients with recent MRI of the brain, but without clinical or radiological signs of elevated ICP, were selected to undergo ocular sonography via three ultrasound planes (inferior, sagittal, and transverse). Optic nerve sheath was measured in each ultrasound view as well with MRI. Image quality scores were assigned for the ultrasound views in different orientations. The three ocular plane views were analyzed for correlation. In addition, correlation of the three ocular ultrasound views with MRI was also performed.
Results:
Correlation analysis showed a wide variability in the correlation between different ultrasound views with magnitude range of 0.1 to 0.8 and directions being both positive and negative. There was a difference in image quality scores between the ultrasound views. The inferior and transverse orientations were superior to the sagittal orientation in achieving high image quality. Comparison to MRI measurements did not demonstrate a significant correlation.
Conclusion:
Our findings suggest that absolute measurements should not be compared across different ultrasound orientations given the wide variability in the correlation between the ultrasound views used to assess the optic nerve sheath. The inferior and transverse ultrasound views are the most likely to yield high-quality images, although the specific view, for the best image, in an individual patient can vary. We would caution against absolute values of ONSD to indicate increased ICP, as it may be view dependent.