Pulse pressure as a function of cuff width.

E A Arcuri1, J L Santos, M Rouch e Silva

  • 1Escola de Enfermagem, Faculdade de Saúde Pública, Universidade de São Paulo, Brasil.

Brazilian Journal of Medical and Biological Research = Revista Brasileira De Pesquisas Medicas E Biologicas
|January 1, 1988
PubMed
Summary

Incorrect cuff width during indirect sphygmomanometry significantly impacts pulse pressure readings. Using a standard cuff often leads to inaccurate blood pressure measurements, underestimating both systolic and diastolic pressures.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Cannabis use and cognitive biases in people with first-episode psychosis and their siblings.

Psychological medicine·2024
Same author

Bacterial bioaugmentation for paracetamol removal from water and sewage sludge. Genomic approaches to elucidate biodegradation pathway.

Journal of hazardous materials·2024
Same author

The contribution of cannabis use to the increased psychosis risk among minority ethnic groups in Europe.

Psychological medicine·2024
Same author

Biochar production from cellulose under reductant atmosphere: influence of the total pyrolysis time.

RSC advances·2023
Same author

Tobacco use in first-episode psychosis, a multinational EU-GEI study.

Psychological medicine·2023
Same author

Novel nonylphenol-degrading bacterial strains isolated from sewage sludge: Application in bioremediation of sludge.

The Science of the total environment·2022

Area of Science:

  • Cardiovascular physiology
  • Medical instrumentation

Background:

  • Accurate arterial pressure measurement is crucial for diagnosing and managing cardiovascular conditions.
  • Indirect sphygmomanometry is a common method, but its accuracy can be affected by technique and equipment.
  • The American Heart Association provides guidelines for arterial pressure measurements, including cuff selection.

Purpose of the Study:

  • To investigate the influence of cuff width on pulse pressure measurements obtained by indirect sphygmomanometry.
  • To compare pulse pressure readings using a standard cuff versus a correctly sized cuff.
  • To evaluate the impact of cuff width on systolic and diastolic arterial pressure estimations.

Main Methods:

  • Indirect sphygmomanometry was performed on 900 subjects.

Related Experiment Videos

  • Two cuff widths were used: a standard 12 cm cuff and a correct cuff (20% wider than arm diameter).
  • Seven cuffs of varying widths (8–14 cm) were employed as correct cuffs, adhering to American Heart Association recommendations.
  • Main Results:

    • Pulse pressure was significantly lower (P < 0.001) with the standard cuff compared to the correct cuff across different diastolic pressure ranges.
    • Mean systolic arterial pressure was underestimated by 12.7 mmHg and mean diastolic arterial pressure by 7.7 mmHg when using the standard cuff.
    • The standard cuff was appropriate for only 50 out of 900 subjects.

    Conclusions:

    • Cuff width is a critical determinant of accuracy in indirect sphygmomanometry.
    • Using a standard cuff when inappropriate can lead to significant over- or underestimation of arterial pressures.
    • Optimal cuff selection, based on arm circumference, is essential for reliable blood pressure monitoring.