A High-Throughput Electrochemiluminescence 7-Plex Assay Simultaneously Screening for Type 1 Diabetes and Multiple Autoimmune Diseases
Published on: May 29, 2020
A high-throughput test for diabetes care: An evaluation of the next generation Roche Cobas c 513 hemoglobin A1C assay
Jessica L Gifford1,2, Trefor Higgins3,4, S M Hossein Sadrzadeh1,2
1Calgary Laboratory Services, 9, 3535 Research Rd NW, Calgary, AB, T2L 2K8, Canada.
Insights
The Roche Cobas c 513 analyzer demonstrates excellent precision and accuracy for glycated hemoglobin A (HbA1C) testing. This automated, high-throughput instrument is suitable for large diagnostic laboratories monitoring diabetes.
Area of Science:
- Clinical chemistry
- Diagnostic instrumentation
- Diabetes diagnostics
Background:
- Glycated hemoglobin A (HbA1C) is the gold standard for diabetes mellitus monitoring.
- Accurate and efficient HbA1C measurement is crucial for effective diabetes management.
- Evaluating new analytical platforms is essential for laboratory efficiency.
Purpose of the Study:
- To assess the analytical performance of the Roche Diagnostics Cobas c 513 analyzer for HbA1C determination.
- To validate the c 513 as a reliable standalone immunoassay analyzer.
Main Methods:
- Analytical performance evaluated for imprecision, accuracy, linearity, and method comparison.
- Specimen stability, interference from hemoglobin variants (Hb variants) and hemoglobin F (HbF), and throughput were assessed.
- Comparison performed against the Roche Cobas Integra 800 CTS system.
Main Results:
- Excellent precision with within-run CVs of 0.5-0.7% and between-run CVs of 0.8-1.3%.
- Good correlation with the Integra system (R=0.998) and minimal bias (-1.6% on average).
- Negligible interference from Hb variants; a negative bias observed with HbF ≥10%. High throughput of 340 samples/hour.
Conclusions:
- The Cobas c 513 is a precise and accurate automated analyzer for HbA1C.
- Its high throughput makes it suitable for large-volume clinical laboratories.
- The analyzer provides reliable results for diabetes monitoring.
Objectives:
The level of glycated hemoglobin A (HbA1C) in blood is the preferred marker for diabetes monitoring and treatment. Here, we evaluate the analytical performance of the Roche Diagnostics Cobas c 513, a stand-alone HbA1C immunoassay analyzer.
Design And Methods:
Performance was assessed with regards to imprecision, accuracy, linearity, method comparison against the Roche Cobas Integra 800 CTS, specimen stability, interference from common hemoglobin variants and hemoglobin F, and throughput.
Results:
Within-run and between-run precisions were 0.5-0.7 and 0.8-1.3%CV, respectively. An average bias of -1.6% to proficiency survey samples was observed. The c 513 correlated well with the Integra (slope = 0.94, y-intercept = 0.50, and correlation coefficient = 0.998). The effect of hemoglobin variants on this assay was negligible while specimens containing ≥10% HbF demonstrated a negative bias. The c 513 instrument can process up to 340 samples per hour.
Conclusions:
The c 513 is a precise, accurate, automated high throughput analyzer for measuring HbA1C in large laboratories.