Related Experiment Video
Updated: Jan 2, 2026

Development of an Electrochemical DNA Biosensor to Detect a Foodborne Pathogen
Published on: June 3, 2018
Electrochemical sensing of lead(II) by differential pulse voltammetry using conductive polypyrrole nanoparticles
Tingting Xu1, Hongqi Dai2, Yongcan Jin2
1Jiangsu Province Key Laboratory of Pulp and Paper Science and Technology, College of Light Industry and Food Engineering, Nanjing Forestry University, Nanjing, 210037, China. salyting@njfu.edu.cn.
Abstract:
An electrochemical sensor for Pb(II) is described that exploits (a) the outstanding adsorption ability of chitosan modified with quaternary ammonium groups (CQAS; cationic) and of lignosulfonate (LSN; anionic), and (b) the good electrical conductivity of polypyrrole nanoparticles (PPy NPs). A glassy carbon electrode (GCE) was modified with PPy NPs and polydopamine, and CQAS and LSN are used as dispersants in PPy. The modified GCE exhibits high selectivity and sensitivity for Pb(II) in the 0.1 to 50 μM concentration range, with a 55 nM detection limit (3σ method). The redox potentials for Pb(II) is around -0.55 V, and the sensor is not interfered by the presence of Hg(II) and Cu(II). The time dependent stability test showed that this sensor can maintain good reproducibility for one month. This sensor was applied to the determination of Pb(II) in wastewater samples. An electrochemical sensor for lead(II) is described that exploits the outstanding adsorption ability of chitosan that carries quaternary ammonium groups (CQAS), of lignosulfonate (LSN), and the good electrical conductivity of polypyrrole nanoparticles. CQAS and LSN are used as dispersants in PPy. The sensor exhibits high selectivity and sensitivity for Pb(II) in the range of 0.1 to 50 μM with a 55 nM detection limit.
More Related Videos
10:48Electrochemical Preparation of Poly3,4-Ethylenedioxythiophene Layers on Gold Microelectrodes for Uric Acid-Sensing Applications
Published on: July 28, 2021
11:18Manufacturing of a Nafion-coated, Reduced Graphene Oxide/Polyaniline Chemiresistive Sensor to Monitor pH in Real-time During Microbial Fermentation
Published on: January 7, 2019
Related Concept Videos
Voltammetric Techniques: Pulse Voltammetry
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential...
Electrodeposition
Electrodeposition can...
Voltammetric Techniques: Linear-Scan (E vs Time)
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...