Related Experiment Video
Updated: Jan 2, 2026

Author Spotlight: Advancing Rapid Detection of Respiratory Pathogens Using Microfluidic Chip
Published on: March 29, 2024
Development of loop-mediated isothermal amplification coupled with nanoparticle-based lateral flow biosensor assay
Yacui Wang1, Yi Wang1, Weiwei Jiao1
1Key Laboratory of Major Diseases in Children, Ministry of Education, National Key Discipline of Pediatrics (Capital Medical University), National Clinical Research Center for Respiratory Diseases, Beijing Key Laboratory of Pediatric Respiratory Infection Diseases, Beijing Pediatric Research Institute, Beijing Children's Hospital, Capital Medical University, Beijing, 100045, China.
Abstract:
Mycoplasma pneumoniae (MP) is one of the most common pathogens causing respiratory tract infection, especially for community-acquired pneumonia (CAP) in school-age children. There was considerable amount of studies on loop-mediated isothermal amplification (LAMP) assay for MP detection. However, the result interpretation of these developed LAMP assays was sophisticated and subjective. Therefore, we developed and evaluated a LAMP coupled with nanoparticle-based lateral flow biosensor (LFB) assay (LAMP-LFB) for simple, reliable, and objective identification of MP (MP-LAMP-LFB). Six primers specific to P1 gene of MP were designed, and the preferred temperature for this assay was confirmed to be 65 °C. The amplification products could be visually interpreted by LFB within 2 min. The MP-LAMP-LFB assay specifically identified DNA templates of MP, and no cross-reactivity with other pathogens was obtained. The limit of the detection for this assay was 600 fg of DNA templates in pure cultures, which was in complete accordance with colorimetric indicator detection and agarose gel electrophoresis analysis. This assay was applied to 209 oropharyngeal swab specimens collected from children with acute respiratory tract infection for clinical evaluation, and compared to real-time PCR detection. Using the LAMP-LFB and real-time PCR assay, the positive rates of MP were 47.8% and 31.6%, respectively. Results suggested that the LAMP-LFB assay displayed high sensitivity compared to real-time PCR method. In summary, LAMP-LFB assay established here was a simple, objective, and sensitive assay for MP detection, which can be widely applied in clinical settings, especially in rural areas.
Insights
A new loop-mediated isothermal amplification coupled with nanoparticle-based lateral flow biosensor (LAMP-LFB) assay offers a simple, objective, and sensitive method for detecting Mycoplasma pneumoniae (MP). This advanced diagnostic tool shows high sensitivity in identifying MP in children with respiratory infections.
Area of Science:
- Microbiology
- Molecular Diagnostics
- Biosensor Technology
Background:
- Mycoplasma pneumoniae (MP) is a primary cause of community-acquired pneumonia (CAP) in children.
- Existing loop-mediated isothermal amplification (LAMP) assays for MP detection often have complex and subjective result interpretations.
- A need exists for a simpler, more objective, and reliable MP detection method.
Purpose of the Study:
- To develop and evaluate a novel LAMP assay integrated with a nanoparticle-based lateral flow biosensor (LAMP-LFB) for straightforward and objective MP identification.
- To assess the diagnostic performance of the MP-LAMP-LFB assay in clinical samples from children with acute respiratory tract infections.
Main Methods:
- Design of six MP-specific primers targeting the P1 gene.
- Optimization of the LAMP reaction at 65°C.
- Utilizing a nanoparticle-based lateral flow biosensor (LFB) for rapid visual detection of amplification products within 2 minutes.
- Clinical evaluation using 209 oropharyngeal swab specimens and comparison with real-time PCR.
Main Results:
- The MP-LAMP-LFB assay demonstrated high specificity, with no cross-reactivity observed against other common pathogens.
- The limit of detection was determined to be 600 fg of MP DNA, consistent with other detection methods.
- Clinical evaluation revealed a higher MP positive rate (47.8%) compared to real-time PCR (31.6%), indicating superior sensitivity.
Conclusions:
- The developed MP-LAMP-LFB assay is a simple, objective, and highly sensitive method for MP detection.
- This assay holds significant potential for widespread clinical application, particularly in resource-limited settings like rural areas.