Technical Note: Performance comparison of ultra-high-resolution scan modes of two clinical computed tomography
Hiroki Kawashima1, Katsuhiro Ichikawa1, Tadanori Takata2
1Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, 920-0942, Japan.
Purpose:
Two ultra-high-resolution (UHR) computed tomography (CT) scanners are clinically available. One is achieved by a CT system with a 0.25 mm × 160 row detector (Detector-UHR), whereas the other is with a 0.6 mm × 32 row detector with in-plane comb filtering in a dual source CT (Comb-UHR). We compared radiation dose efficiencies of the two UHR modes to that of a routine scan mode (RS), using physical image quality measures for an assumed condition of abdominal CT angiography (CTA).
Methods:
A wire phantom, a 300-mm cylindrical water bath phantom, and a microdisk phantom were used for measuring the modulation transfer function (MTF), noise power spectrum (NPS), and slice sensitivity profiles (SSP), respectively, at CTDIvol of 20 mGy. Images with minimal slice thicknesses were reconstructed by filtered back projection methods. System performance functions (SPF2 ) based on the prewhitening theorem were calculated by dividing MTF2 by NPS measurements. The ideal observer's detectability index (d'2 ) was also estimated for a task corresponding to a 1-mm diameter vessel. Furthermore, a bar-pattern phantom placed in a water phantom resembling an adult abdomen was scanned, and the visibility of the bars was observed.
Results:
System performance function (SPF2 ) results showed that Comb-UHR has a 70% dose efficiency compared to RS and provides better than twofold SPF2 compared to Detector-UHR. The d'2 values of Detector-UHR, Comb-UHR, and RS were 6.5, 14.7, and 16.0, respectively. Although the bar-pattern phantom images were consistent with the SPF results, bar widths < 1.0 mm for Detector-UHR and < 0.75 mm for Comb-UHR and RS were not resolved.
Conclusions:
Though both the UHR modes exhibited system performances extending to 2.6 cycles/mm, they appeared not to be advantageous compared to RS in the conditions applicable to abdominal CTA, because of their insufficient dose efficiencies.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT