Comparison between His-bundle pacing guided by Ensite NavX system and conventional fluoroscopy
Yiran Hu1, Ligang Ding1, Wei Hua2
1State Key Laboratory of Cardiovascular Disease, Arrhythmia Center, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Ensite NavX system significantly reduces fluoroscopy exposure during His-bundle pacing (HBP). This novel approach is effective, showing comparable pacing parameters to conventional methods with substantially lower radiation doses.
Area of Science:
- Cardiology
- Medical Devices
- Electrophysiology
Background:
- Conventional fluoroscopy guidance for His-bundle pacing (HBP) is associated with high fluoroscopy exposure.
- Minimizing radiation dose is crucial for patient safety during cardiac procedures.
Purpose of the Study:
- To evaluate the feasibility of using the Ensite NavX system for His-bundle pacing.
- To compare pacing parameters and fluoroscopy doses between the Ensite NavX system and conventional fluoroscopy guidance.
Main Methods:
- A prospective study enrolled 30 patients undergoing HBP, with 15 patients in the Ensite NavX group and 15 in the conventional fluoroscopy group.
- Procedure time, fluoroscopy doses (FD), and pacing parameters were recorded at implant and 3-month follow-up.
Main Results:
- Successful HBP was achieved in 93.3% of patients using the Ensite NavX system.
- The Ensite NavX group had significantly lower total FD (1.4 ± 0.5 mGy) and His lead FD (0.4 ± 0.3 mGy) compared to the conventional group (16.1 ± 2.2 mGy and 12.5 ± 2.2 mGy, respectively).
- No significant differences in pacing parameters were observed between the groups at implant and follow-up.
Conclusions:
- His-bundle pacing guided by the Ensite NavX system is an efficient procedure.
- The Ensite NavX system offers a significant reduction in fluoroscopy exposure compared to conventional fluoroscopy.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Endoscopic Procedures III: Video Capsule Endoscopy
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...