SUV variability in EARL-accredited conventional and digital PET
Daniëlle Koopman1,2, Pieter L Jager3, Cornelis H Slump4
1Department of Nuclear Medicine, Isala, Dokter van Heesweg 2, 8025, AB, Zwolle, the Netherlands. d.koopman@isala.nl.
EJNMMI Research
|December 12, 2019
Summary
This study shows that conventional and digital PET scanners have limited SUV variability (up to 8%) when using EARL standardization. These findings confirm the effectiveness of EARL standardization for consistent PET imaging across different scanners.
Area of Science:
- Nuclear Medicine
- Medical Imaging
- Oncology
Background:
- Standardization of SUV measurements is critical for multi-center PET studies.
- Evaluating SUV variability between conventional and digital PET scanners is essential for reliable cancer imaging.
- This study compares SUV reproducibility using EARL-accredited scanners.
Purpose of the Study:
- To assess SUV variability in whole-body FDG-PET scans between conventional (cPET) and digital (dPET) scanners.
- To compare SUVmean, SUVpeak, and SUVmax differences using EARL-compatible reconstructions.
- To evaluate the impact of scanner technology on SUV measurements in cancer patients.
Main Methods:
- Prospective study comparing 50 patients' FDG-PET scans from cPET and dPET scanners.
- Utilized EARL-accredited protocols and reconstructions for both scanner types.
- Measured SUVmean, SUVpeak, SUVmax, and lesion diameter in up to 5 lesions per patient.
- Calculated relative SUV differences (ΔSUV) and repeatability coefficients.
Main Results:
- Included 128 lesions (average size 19 ± 14 mm).
- Average ΔSUV ranged from 6-8%, with higher values on dPET (p < 0.001).
- ΔSUVmax was significantly higher than ΔSUVmean and ΔSUVpeak (8% vs. 6-7%, p < 0.05).
- Repeatability coefficients were 27% for ΔSUVmean/peak and 33% for ΔSUVmax (p < 0.001).
Conclusions:
- Limited SUV variability (up to 8%) was observed between conventional and digital PET scanners with EARL accreditation.
- Few lesions exhibited SUV differences exceeding 30%, supporting scanner comparability.
- EARL standardization effectively ensures consistent SUV measurements across different PET scanner technologies.
Related Concept Videos
Feedback control systems
645
Feedback control systems are categorized in various ways based on their design, analysis, and signal types.
Linear feedback systems are theoretical models that simplify analysis and design. These systems operate under the principle that their output is directly proportional to their input within certain ranges. For instance, an amplifier in a control system behaves linearly as long as the input signal remains within a specific range. However, most physical systems exhibit inherent nonlinearity...
Linear feedback systems are theoretical models that simplify analysis and design. These systems operate under the principle that their output is directly proportional to their input within certain ranges. For instance, an amplifier in a control system behaves linearly as long as the input signal remains within a specific range. However, most physical systems exhibit inherent nonlinearity...
645
Electronic Distance Measuring Instruments
408
Electronic Distance Measuring Instruments (EDMs) are essential tools in modern surveying, offering precise distance measurements by emitting electromagnetic signals and calculating the time required for these signals to travel to a target and return. Two primary types of signals are used in EDMs — light waves and microwaves — each suited to specific environmental and distance requirements. Light-wave-based EDMs utilize either infrared or laser light, providing high accuracy over...
408