Using simulation to explore the impact of device design on the learning and performance of peripheral intravenous
Bronwyn Reid-McDermott1,2, Maryanne Browne3,2, Dara Byrne1
11Irish Centre for Applied Patient Safety and Simulation, School of Medicine, National University of Ireland Galway, Galway, Ireland.
Background:
The design of medical devices impacts upon the performance of healthcare professionals and patient safety. However, multiple devices serving the same function are often available. The purpose of this study was to use simulation as a means of examining the impact of differences in device design on (1) learning of, or attainment of behavioral fluency in, peripheral intravenous cannulation (PIVC); and (2) the generalization, or transfer, of learning on one device to performance of PIVC using an untrained device.
Methods:
A total of 25 final cycle medical students participated in this study which used a randomized two-group design. Participants were randomly assigned to learn PIVC using either a closed PIVC device (a single device which consists of an intravenous cannula with a pre-attached extension tube; n = 14) or an open PIVC device (a two-piece device made up of an intravenous cannula and a separate extension tube which is attached following insertion of the cannula; n = 11). Task analyses were developed for the performance of PIVC using each device. Subsequently, simulation-based fluency training was delivered to both groups using their assigned PIVC device, and continued for each participant until the fluency criterion was achieved. Following achievement of fluency, participants were asked to perform PIVC using the untrained device (i.e., the PIVC device that they had not been trained on).
Results:
All participants in both groups met the fluency criterion, and no significant differences were observed in the number of trials or total training required by groups to achieve fluency. Participants in both groups improved significantly from baseline (M = 11.69) to final training trial (M = 100). However, a significant decrement in performance (M = 81.5) was observed when participants were required to perform PIVC using the untrained device.
Conclusions:
Participants achieved fluency in PIVC regardless of the device used. However, significant decrements in performance were observed when participants were required to perform PIVC using a novel device. This finding supports the need for careful consideration of devices purchased and supplied in the clinical setting, and the need for training prior to the introduction of novel devices or for new staff members.