Comments on "A novel high input impedance front-end for capacitive biopotential measurement"
1Department of Electronic Engineering, Universitat Politècnica de Catalunya (BarcelonaTech), Barcelona, Catalunya, Spain.
Abstract:
A front-end for biopotential sensing in wearable medical devices has been recently proposed which is claimed to provide 100 GΩ input impedance by manually matching two resistor pairs in a positive- and a negative-feedback loop around an operational amplifier (op amp); the cost being that the equivalent input noise voltage doubles with respect to a simple non-inverting amplifier. The ECG acquired with capacitive (sic) electrodes through a cotton shirt is presented as a proof of the performance of the proposed circuit. It turns out, however, that the analysis ignores op amp's input capacitance hence the effort to achieve a very high input resistance seems futile. Further, cotton is highly hygroscopic hence not an appropriate dielectric, so that there is no proof that the electrodes tested were actually capacitive. This comment addresses these two problems and some additional conceptual and methodological inaccuracies found in the paper.
Related Concept Videos
Bode Plots Construction
Potentiometry: Membrane Electrodes
Equivalent Capacitance
The following strategies are adopted to calculate...
Equivalent Capacitance
Design Example: Capacitance Multiplier Circuit
The circuit illustrated in Figure 1 below incorporates two op-amps, with the first operating as a voltage follower and the second acting as an inverting amplifier.