In Search of a Simple and Reliable Method of Measuring Intra-abdominal Pressure in Peritoneal Dialysis Patients
Tharshika Thangarasa1, Dana Foisy2, Julie Leidecker3
1Faculty of Psychiatry, University of Toronto, Toronto, ON, Canada.
Background:
Patients treated with peritoneal dialysis (PD) are at increased risk of developing mechanical complications such as dialysate leaks and hernias thought to be partially related to an increase in intra-abdominal pressure (IAP) secondary to dialysate in the abdomen. However, measurement of IAP requires specialized equipment that is not readily available in the home dialysis unit.
Objectives:
To develop a reliable method of measuring IAP in PD patients that could be easily used in the home dialysis unit. We hypothesized that the handheld Stryker pressure monitor would be suitable for this purpose via connection to the PD catheter.
Design:
Cross-sectional.
Setting:
Tertiary Care Hospital, Ottawa, Ontario, Canada.
Patients:
Patients who were having a PD catheter inserted via laparoscopic surgery at The Ottawa Hospital were recruited for the study.
Measurements:
With the patients at end-expiration, the IAP measured with the Stryker monitor connected to the PD catheter was compared with the insufflator pressures of 15, 10, and 5 mm Hg.
Methods:
Bland-Altman plots were constructed and intraclass correlation coefficients were calculated for each pressure.
Results:
Twelve patients participated in the study: 9 men and 3 women. They were on average 53 ± 15 years old and 81 ± 13.4 kg. Two patients had to be excluded from the analysis due to difficulties zeroing the Stryker pressure monitor at the time of surgery. There were also rapid fluctuations in the insufflator pressure recording, creating additional challenges in comparing the 2 measurements at end-expiration. The 95% limits of agreement for the Bland-Altman plots ranged from 7.9 (@15 mm Hg) to 12.2 (@10 mm Hg). The intraclass correlation coefficients for reliability of the individual measurements ranged from 0.015 (10 mm Hg) to 0.634 (15 mm Hg).
Limitations:
Small sample size and lack of a gold standard comparator may have affected our results.
Conclusions:
In our study, we used the operating room insufflator as the gold standard for measuring IAP. By Bland-Altman plots and intraclass correlation coefficients, the pressure values obtained with the Stryker pressure monitor were not a reliable estimate of insufflator IAP especially at lower pressures. Further studies are needed to identify an ideal tool for measurement of IAP to guide PD management.
Related Concept Videos
Peritoneal Dialysis I: Introduction and Procedure
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Peritoneal Dialysis III: Nursing Management
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...