Background:
Bronchial occlusion using an endobronchial Watanabe spigot (EWS) is reportedly effective for intractable bronchopleural fistula. Here, we describe a rapid and easy method for bronchial occlusion using a guide sheath (GS) and curette.
Methods:
Thirty consecutive patients who underwent bronchial occlusion under mild sedation between October 2014 and February 2018 were enrolled. The devices used were a flexible bronchoscope (BF-1T260 or BF-1TQ290), GS (SG-201C; with 30 mm of the proximal end cutaway), and a CC-4CR-1 curette (all supplied by Olympus Ltd). The curette was inserted into the GS with the tip of the curette exposed outside the GS. The curette and GS were inserted into the bronchoscope. The EWS attached to the curette tip was inserted into the target bronchus and left in position by pulling the curette back through the GS while pushing the EWS with the GS under the bronchoscopic view. The success rate and procedure time were recorded.
Results:
Bronchial occlusion with an EWS was performed on 143 target bronchi (2 to 9 bronchi/patient). The bronchial occlusion success rate was 98.6%. The median procedure time for bronchial occlusion per EWS on video recordings of the 10 most recent procedures was 110 (range, 40 to 521) seconds. The target bronchial occlusion success rate was 100%. This method enabled easy insertion of the EWS, even in the sharply branching upper lobe bronchus. No complications were observed.
Conclusion:
Bronchial occlusion using a GS and curette is a rapid and easy technique even in a sharply branching target bronchus.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...