Reduction of Occupational Exposure Using a Novel Tungsten-Containing Rubber Shield in Interventional Radiology
Kenta Kijima1, Anchali Krisanachinda2, Mikoto Tamura1
1Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, Osaka, Japan.
Abstract:
This study investigates whether a novel tungsten-containing rubber shield could be used as substitute shielding material in interventional radiology to reduce the occupational exposure of operators to scattered radiation from a patient. The tungsten-containing rubber is a lead-free radiation-shielding material that contains as much as 90% tungsten powder by weight. Air kerma rates of scattered radiation from solid-plate phantoms, simulating a patient, were measured with a semiconductor dosimeter at the height of the operator's eye (1,600 mm from the floor), chest (1,300 mm), waist (1,000 mm), and knee (600 mm) with and without tungsten-containing rubber shielding (1-5 mm thickness). The tungsten-containing rubber and a commercial shielding material (RADPAD) were affixed onto the phantom on the operator's side, and reductions in air kerma rates were compared. Reduction rates for tungsten-containing rubber shielding with thicknesses of 1, 2, 3, 4, and 5 mm at each height level were as follows: 70.37 ± 0.40%, 72.17 ± 0.29%, 72.95 ± 0.31%, 72.58 ± 0.35%, and 73.63 ± 0.63% at eye level; 76.36 ± 0.19%, 77.13 ± 0.10%, 77.36 ± 0.14%, 77.62 ± 0.25%, and 77.66 ± 0.14% at chest level; 67.78 ± 0.31%, 68.12 ± 0.19%, 68.88 ± 0.28%, 68.97 ± 0.14%, and 68.85 ± 0.45% at waist level; and 0.14 ± 0.94%, 0.72 ± 0.56%, 1.08 ± 0.74%, 1.77 ± 0.80%, and 1.79 ± 1.82% at knee level, respectively. Reduction rates with RADPAD were 61.80 ± 0.67%, 60.33 ± 0.61%, 64.70 ± 0.25%, and 0.14 ± 0.66% at eye, chest, waist, and knee levels, respectively. The shielding ability of the 1 mm tungsten-containing rubber was superior to that of RADPAD. The tungsten-containing rubber could be employed to minimize an operator's radiation exposure instead of the commercial shielding material in interventional radiology.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Nuclear Power
Nuclear Fuels
Nuclear fuel consists of a fissile isotope, such as uranium-235, which must be present in sufficient quantity to provide a...