Summary
Orthogonal biplanar X-rays analyzed with graphical analysis (GA), ray tracing (RT), and perspective projection (PP) enable precise stereotactic intraoperative localization (StIL) for deep brain stimulation (DBS) electrode placement.
Area of Science:
- Neurosurgery
- Medical Imaging
- Biomedical Engineering
Background:
- Deep brain stimulation (DBS) efficacy relies on accurate electrode placement.
- Current stereotactic intraoperative localization (StIL) methods include fluoroscopy, CT, and MRI.
- The utility of biplanar X-ray for StIL requires detailed evaluation.
Purpose of the Study:
- To assess the feasibility of using orthogonal biplanar X-rays with graphical analysis (GA), ray tracing (RT), and perspective projection (PP) for StIL.
- To compare the accuracy of these X-ray based methods against postoperative CT scans.
Main Methods:
- A retrospective review compared electrode tip positions from postoperative CT scans with intraoperative biplanar X-ray methods.
- Orthogonal biplanar X-rays utilized referential spheres and pins for localization.
- Analysis included graphical analysis (GA), ray tracing (RT), and perspective projection (PP).
Main Results:
- Mean Euclidean errors for DBS electrode position were: GA (1.58 mm ±0.75), RT (0.74 mm ±0.45), and PP (1.07 mm ±0.64).
- GA accuracy improved with ventriculograms; RT and PP correlated with third ventricular structures.
- RT offered stability, PP provided flexibility, and O-arm imaging showed sub-millimeter errors.
Conclusions:
- Graphical analysis, ray tracing, and perspective projection are viable for precise StIL in DBS using biplanar X-ray.
- These techniques can be adapted for other stereotactic procedures and biplanar imaging applications.
- Potential applications include angiograms, radiosurgery, and injection therapeutics.