Related Experiment Video
Updated: Jan 1, 2026

Using an Automated Hirschberg Test App to Evaluate Ocular Alignment
Published on: March 24, 2020
[Invention of Optical Sight in Mobile Radiography with Anti-scatter Grid]
1Department of Radiology, Kitakyusyu Tsuyazaki Hospital (Current address: Department of Radiology, Kitakyusyu Munakata Cyuuou Hospital).
Abstract:
In radiography with anti-scatter grid, it is important to make sure that the X-ray beam direct exactly perpendicular to the grid plane. However, it is so difficult to ensure in mobile radiography. An optical sight to ensure X-ray alignment in mobile radiography with anti-scatter grid was devised. The device measures the X-ray beam angle respect to the grid plane by utilizing collimator-lamp. Computed radiography of water phantom on inclined bedding with anti-scatter grid (6 : 1) were done by aid of devised optical sight 20 times. The result showed that the average alignment error of the radiographies by aid of devised optical sight was within 1°, and the maximum error was<2°.

