Related Experiment Video

Updated: Jan 1, 2026

Smartphone Fundus Photography
05:51

Smartphone Fundus Photography

Published on: July 6, 2017

40.0K

Non-contact smartphone-based fundus imaging compared to conventional fundus imaging: a low-cost alternative for

Maximilian W M Wintergerst1, Michael Petrak1, Jeany Q Li1

  • 1Department of Ophthalmology, University of Bonn, Ernst-Abbe-Str. 2, 53127, Bonn, Germany.

Scientific Reports
|December 25, 2019
PubMed

Insights

Smartphone-based fundus imaging (SBFI) offers an affordable, mobile alternative for retinopathy of prematurity (ROP) screening. This low-cost method shows potential for improving ROP care in underserved regions.

Area of Science:

  • Ophthalmology
  • Medical Technology
  • Public Health

Background:

  • Retinopathy of prematurity (ROP) is a leading cause of treatable childhood blindness.
  • Current telemedicine screening for ROP relies on expensive equipment, limiting access in low-resource settings.
  • Smartphone-based fundus imaging (SBFI) presents a potential solution for cost-effective, mobile ROP screening.

Purpose of the Study:

  • To compare the feasibility of non-contact SBFI with conventional contact fundus imaging (CFI) for ROP screening and documentation.
  • To evaluate the diagnostic accuracy of SBFI for ROP and associated "plus disease" compared to CFI.

Main Methods:

  • Twenty-six eyes were imaged using both non-contact SBFI and conventional contact fundus imaging (CFI).
  • Examination time, field-of-view, and level of detail were compared between SBFI and CFI.
  • Agreement with clinical evaluation, sensitivity, and specificity for ROP and plus disease detection were assessed for both imaging methods.

Main Results:

  • SBFI had a smaller field-of-view (1:2.5 ratio) and longer examination time (109.0 ± 57.8s vs. 75.9 ± 36.3s, p < 0.01) than CFI, but offered equal detail.
  • Both SBFI and CFI demonstrated good agreement with clinical evaluations for ROP stage and plus disease detection.
  • High sensitivity and specificity for detecting ROP and plus disease were achieved with both methods, with SBFI showing 90%/100% and 88%/93% for plus disease and ROP, respectively.

Conclusions:

  • Smartphone-based fundus imaging (SBFI) is a viable, non-contact, low-cost alternative to CFI for ROP screening and documentation.
  • SBFI has the potential to significantly enhance ROP care accessibility in middle- and low-resource settings.
  • This technology could broaden access to essential eye care for premature infants in underserved populations.