Single-frame-based column fixed-pattern noise correction in an uncooled infrared imaging system based on weighted

Applied Optics
|December 25, 2019
PubMed
Summary

This study presents a new single-frame algorithm to remove column fixed-pattern noise (FPN) from uncooled infrared images. The novel non-uniformity correction (NUC) method effectively eliminates stripe artifacts, improving image clarity.

Related Concept Videos

Calibration Curves: Linear Least Squares01:20

Calibration Curves: Linear Least Squares

A calibration curve is a plot of the instrument's response against a series of known concentrations of a substance. This curve is used to set the instrument response levels, using the substance and its concentrations as standards. Alternatively, or additionally, an equation is fitted to the calibration curve plot and subsequently used to calculate the unknown concentrations of other samples reliably.
For data that follow a straight line, the standard method for fitting is the linear...
4.0K
Infrared (IR) Spectroscopy: Overview01:09

Infrared (IR) Spectroscopy: Overview

When electromagnetic radiation passes through a material, atoms or molecules transition from a lower to a higher energy state by absorbing radiation corresponding to the energy difference between the two states. The absorption of infrared (IR) radiation causes transitions between vibrational energy levels in a molecule. Therefore, IR spectroscopy is a useful analytical tool for determining the molecular structure of molecules.
Different compounds display unique properties due to their...
4.5K
IR Spectrometers01:25

IR Spectrometers

There are two main infrared (IR) spectrophotometers: dispersive IR spectrometers and Fourier transform infrared (FTIR) spectrometers. In a dispersive IR spectrometer, a beam of infrared radiation produced by a hot wire is divided into two parallel equal-intensity beams using mirrors. One beam passes through the sample, while another is a reference beam. The beams then move through the monochromator, which separates the radiations into a continuous spectrum of different frequencies. The...
2.1K
Root Mean Square00:57

Root Mean Square

If in an experiment, data values have a probability of being both positive and negative, neither the arithmetic mean, the geometric mean, nor the harmonic mean can be used to calculate the central tendency of the data set. In particular, if the positive and negative values are equally likely, the arithmetic mean is close to zero.
For example, consider the velocity of gas molecules in a container. The gas molecules are moving in different directions, which might impart positive and negative...
3.6K
IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations01:08

IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations

Identical bonds within a polyatomic group can stretch symmetrically (in-phase) or asymmetrically (out-of-phase). Similar to hydrogen bonding, these vibrations also influence the shape of the IR peak. Generally, asymmetric stretching frequencies are higher than symmetric stretching frequencies. For example, primary amines exhibit two distinct IR peaks between 3300–3500 cm−1 corresponding to the symmetric and asymmetric N-H stretching, while secondary amines exhibit a single...
1.7K
IR Spectroscopy: Hooke's Law Approximation of Molecular Vibration01:16

IR Spectroscopy: Hooke's Law Approximation of Molecular Vibration

A covalently bonded heteronuclear diatomic molecule can be modeled as two vibrating masses connected by a spring. The vibrational frequency of the bond can be expressed using an equation derived from Hooke's law, which describes how the force applied to stretch or compress a spring is proportional to the displacement of the spring. In this case, the atoms behave like masses, and the bond acts like a spring.
According to Hooke's law, the vibrational frequency is directly proportional to...
2.7K