Lesion Revascularisation Subsequent to Femoropopliteal Spot Stenting Using the Multi-LOC Stent Delivery System
Martin Sigl1, Klaus Amendt2, Matthias Waliszewski3,4
1First Department of Medicine, Universitätsmedizin Mannheim (UMM), Mannheim, Germany martin.sigl@umm.de.
Background/Aim:
Restenosis and stent fractures are well-characterised treatment failures following femoropopliteal (FP) stent-based interventions. We aimed to determine patterns of restenosis and fracture rates following focal stenting of FP arteries.
Patients And Methods:
This retrospective study investigated angiographic patterns of restenosis and the occurrence of stent fractures following focal FP stenting with the multiple stent delivery system VascuFlex® Multi-LOC.
Results:
We identified 10 patients with 10 (of 129) FP lesions (mean length 11.7±4.6 cm) and a total of 51 (of 646) Multi-LOC stents that underwent clinically driven target lesion revascularizations (TLR) after 11.5±9.2 months, due to symptomatic recurrence of stenosis. None of the Multi-LOC stents had fractured. No isolated in-stent restenosis (>50%) was observed. Median diameter lumen loss was significantly more pronounced at the inter-stent segments (27.0%) compared to in-stent segments (7.8%, p=0.023).
Conclusion:
No evidence of fracture or susceptibility to stent-related restenosis using Multi-LOC stents was observed. Focal FP stenting may be more effective when combined with strategies known to reduce restenosis in non-stented artery segments.