Comparisons of Effective Fields of Two Ultra-Widefield Ophthalmoscopes, Optos 200Tx and Clarus 500
Yoshitsugu Matsui1, Atsushi Ichio1, Asako Sugawara1
1Department of Ophthalmology, Mie University Graduate School of Medicine, Tsu, Japan.
Purpose:
To compare the effective fields of the Optos 200Tx® and Clarus 500™, two ultra-widefield ophthalmoscopes, based on their ability to image branches of retinal vessel in the four retinal quadrants.
Methods:
Ninety retinal images from 90 patients with various eye diseases were studied. All patients had undergone 200° retinal imaging to obtain a single image of Optos (O) and the montage of two images of the Clarus (C). The highest number of traceable vessel branches in the four retinal quadrants was determined by two masked raters. An image was classified as "O > C" when the number of identifiable branch was greater in the Optos than the Clarus, as "O = C" when the number was equal and as "O < C" when the number was fewer in the Optos than the Clarus.
Results:
The appearance probability of "O > C" was significantly higher at the upper temporal quadrant than "O < C" (p < 0.01 for both raters). In contrast, the appearance probability of "O < C" was significantly higher at the lower nasal quadrant than "O > C" (p < 0.01 for both raters). There were no significant differences in the appearance probability between "O > C" and "O < C" at the other two retinal quadrants (p > 0.50 for both raters).
Conclusions:
These results demonstrate that the effective field of views was different between the two devices at different retina quadrants. Further studies are needed to clarify possible factors such as artifacts by the eyelashes, differences in the depth of focus, motion of the device, and different locations of the images on the effective field of views.
More Related Videos
05:14Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
06:19In Vivo Imaging of Cx3cr1gfp/gfp Reporter Mice with Spectral-domain Optical Coherence Tomography and Scanning Laser Ophthalmoscopy
Published on: November 11, 2017
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...