Glucose Variability Assessed with Continuous Glucose Monitoring: Reliability, Reference Values, and Correlations with
Yuri D Foreman1,2, Martijn C G J Brouwers2,3, Carla J H van der Kallen1,2
1Department of Internal Medicine, Maastricht University Medical Center+, Maastricht, the Netherlands.
Reliable glucose variability (GV) measurement using continuous glucose monitoring (CGM) requires at least three days. Established reference ranges and correlations with oral glucose tolerance tests (OGTT) aid GV assessment when CGM is unavailable.
Area of Science:
- Endocrinology
- Metabolic Health
- Diabetes Management
Background:
- Glucose variability (GV) is a key glycemic control marker, but its measurement needs methodological refinement.
- Continuous glucose monitoring (CGM) offers insights into GV, yet optimal assessment duration and reference values require clarification.
Purpose of the Study:
- To determine the minimum number of monitoring days for reliable GV assessment using CGM.
- To establish reference values for CGM-derived GV indices.
- To correlate GV metrics with established glycemic control indicators like HbA1c and oral glucose tolerance test (OGTT) results.
Main Methods:
- Analysis of cross-sectional data from The Maastricht Study (n=851) with over 48 hours of CGM data.
- Calculation of reliability using the Spearman-Brown prophecy formula.
- Determination of reference ranges (2.5th-97.5th percentiles) and correlations (Spearman's rho) for CGM indices (mean sensor glucose, standard deviation, coefficient of variation).
Main Results:
- Reliable GV measurement was achieved with two monitoring days for mean sensor glucose (MSG) and standard deviation (SD), and three days for coefficient of variation (CV).
- Reference ranges in individuals with normal glucose metabolism were established: MSG (90.5-120.6 mg/dL), SD (7.9-24.8 mg/dL), and CV (7.74%-22.45%).
- MSG correlated strongest with fasting plasma glucose; SD and CV showed significant correlations with 1-hour OGTT values and incremental glucose peak (IGP).
Conclusions:
- The findings provide essential reliability data and reference values for CGM-based GV studies.
- One-hour OGTT and IGP values serve as viable surrogate markers for GV when CGM data is inaccessible.
More Related Videos
Related Concept Videos
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...
Hormones Regulating Blood Glucose
In addition to accelerating glucose uptake and utilization, insulin has...
Glucose Homeostasis: Pancreatic Islets and Insulin Secretion
Insulin and C-peptide are...
Glucose Absorption Into the Small Intestine