Effect of the valve design on pressure drop, pressure recovery, and spatial positioning of vena contracta
Related Experiment Video
Updated: Dec 31, 2025

Design of a Cyclic Pressure Bioreactor for the Ex Vivo Study of Aortic Heart Valves
Published on: August 23, 2011
Giordano Tasca1,2, Federico Lucherini2,3, Claudia Romagnoni3,4
1Cardiac Surgery Unit, Heart Health Center, King Saud Medical City, Riyadh, Kingdom of Saudi Arabia.
Background:
Bioprostheses are complex structures and yield a very complex fluid dynamics. Hence, it can be hypothesized that prosthesis structural characteristics affect the position of the vena contracta and, consequently, influences the pattern and the extent of pressure recovery downstream from the vena contracta.
Materials And Methods:
The study was performed on pericardial aortic prostheses, specifically Crown 21 and 23 (LivaNova PLC, UK), Trifecta 19 and 21 (Edwards Lifescience, USA), and Magna 19 and 21(Abbott, USA), tested in an "ad hoc" devised steady flow loop circuit at four flow rates (10, 15, 20, and 25 L/min). Fluid dynamic quantities were obtained by direct pressure measurement and Doppler interrogation.
Results:
Pressure drop at 25 L/min flow rate was 26.5 ± 0.3 mm Hg and 14.9 ± 0.1 mm Hg for the Trifecta 19 and 21, 37.1 ± 1.0 mm Hg and 27.3 ± 0.4 mm Hg for the Magna 19 and 21, and 36.6 ± 1.0 mm Hg and 22.7 ± 0.1 mm Hg for Crown 21 and 23, respectively. The vena contracta was shorter for Trifecta compared with the Magna and the Crown in which it developed further downstream and as far as 1 cm from the valve leaflets fringes. The pressure recovery was 54% ± 1% for Trifecta 21, 39% ± 1% for Magna 21, and 41% ± 2% for Crown 23 with different patterns.
Conclusion:
The design of bioprosthesis affects pressure recovery and the position of the vena contracta. The different patterns of pressure recovery might have clinical impact.
Related Concept Videos
Venous Return
What is Venous Return?
Venous return refers to the rate at which blood flows back to the heart from the body's peripheral veins. It's an integral part of the circulatory system...
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Free Jet
Blood Flow
Veins as Blood Reservoirs
Regulation of Stroke Volume
Preload refers to the degree of stretch on the heart before it contracts. It's analogous to the stretching of a rubber band; the more it's stretched, the more forcefully it snaps back. This concept is encapsulated in the Frank-Starling law of the...