Related Experiment Video
Updated: Dec 31, 2025

Assessment of Right Ventricular Structure and Function in Mouse Model of Pulmonary Artery Constriction by Transthoracic Echocardiography
Published on: February 3, 2014
A Comparison of Transesophageal to Transthoracic Echocardiographic Measures of Right Ventricular Function
Shayne Michael Roberts1, John Klick1, Adrian Fischl1
1Department of Anesthesiology and Perioperative Medicine, Penn State Health Milton S Hershey Medical Center, Hershey, PA.
Objectives:
To assess the concordance between transesophageal echocardiographic (TEE) and transthoracic echocardiograpic (TTE) measures of right ventricular (RV) function using standard 2-dimensional and Doppler methods. The authors hypothesized that there would be significant disagreement in tricuspid annular plane systolic excursion (TAPSE), fractional area change, right-sided index of myocardial performance, and tricuspid annular systolic velocity (S').
Design:
Prospective observational.
Setting:
Cardiac operating room at a single academic medical center.
Participants:
All adult patients undergoing elective cardiac surgery at a single tertiary care academic medical center over 6 months.
Interventions:
None.
Measurements And Main Results:
The fractional area change, S', TAPSE, right-sided index of myocardial performance, and tricuspid annular diameter were measured with TEE and TTE to assess for concordance using the concordance correlation coefficient and paired t tests, including 95% confidence limits. The study demonstrated that quantitative measures of RV function by TEE correlate poorly with TTE measurements in close temporal proximity under similar hemodynamic conditions.
Conclusions:
When performing an assessment of RV function, transesophageal echocardiographers should exercise caution when extrapolating data validated by TTE to TEE studies. Measures of RV function by TEE tend to have fair agreement to TTE measurements obtained in close temporal proximity under similar hemodynamic conditions. Most importantly, the present study showed that TAPSE and S' values obtained from the modified transgastric RV inflow view tend to have lower values than those measured with TTE. Given the propensity for underestimating measurements from the modified transgastric RV inflow view, the authors conclude that values equal to or greater than established norms for tricuspid annular motion may be used to establish normal-but not abnormal-RV function.
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Rheumatic Heart Disease II: Clinical Manifestations and Diagnostic Studies
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...