Influence of Socket-shield technique on the biomechanical response of dental implant: three-dimensional finite
João Paulo Mendes Tribst1, Amanda Maria de Oliveira Dal Piva1, Alexandre Luiz Souto Borges1
1Department of Dental Materials and Prosthodontics, Institute of Science and Technology, São Paulo State University (UNESP), São José dos Campos, Brazil.
Abstract:
This study evaluated the bone microstrain, displacement, and stress distribution according to the surgical technique (conventional or Socket-shield) and evaluation period (immediately after implant installation or after healing). Each condition was modeled for the finite element analysis, totaling four groups, with a morse-taper implant and a cemented prosthesis. The maximum displacement, von Mises stress, and bone microstrain yielded higher values during the immediate stage, without a difference between Socket-shield and conventional treatments. The use of the Socket-shield technique does not negatively impact the biomechanical behavior of an implant-supported prosthesis immediately after healing from the implant installation.