Intraoperative computed tomography imaging for dose calculation in intraoperative electron radiation therapy: Initial
Verónica García-Vázquez1, Felipe A Calvo1,2,3,4, María J Ledesma-Carbayo5,6
1Instituto de Investigación Sanitaria Gregorio Marañón, Madrid, Comunidad de Madrid, Spain.
Abstract:
In intraoperative electron radiation therapy (IOERT) the energy of the electron beam is selected under the conventional assumption of water-equivalent tissues at the applicator end. However, the treatment field can deviate from the theoretic flat irradiation surface, thus altering dose profiles. This patient-based study explored the feasibility of acquiring intraoperative computed tomography (CT) studies for calculating three-dimensional dose distributions with two factors not included in the conventional assumption, namely the air gap from the applicator end to the irradiation surface and tissue heterogeneity. In addition, dose distributions under the conventional assumption and from preoperative CT studies (both also updated with intraoperative data) were calculated to explore whether there are other alternatives to intraoperative CT studies that can provide similar dose distributions. The IOERT protocol was modified to incorporate the acquisition of intraoperative CT studies before radiation delivery in six patients. Three studies were not valid to calculate dose distributions due to the presence of metal artefacts. For the remaining three cases, the average gamma pass rates between the doses calculated from intraoperative CT studies and those obtained assuming water-equivalent tissues or from preoperative CT studies were 73.4% and 74.0% respectively. The agreement increased when the air gap was included in the conventional assumption (98.1%) or in the preoperative CT images (98.4%). Therefore, this factor was the one mostly influencing the dose distributions of this study. Our experience has shown that intraoperative CT studies are not recommended when the procedure includes the use of shielding discs or surgical retractors unless metal artefacts are removed. IOERT dose distributions calculated under the conventional assumption or from preoperative CT studies may be inaccurate unless the air gap (which depends on the surface irregularities of the irradiated volume and on the applicator pose) is included in the calculations.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT