Small-caliber chest tubes, often used for pneumothorax, have high complication rates. A root cause analysis identified device design and human error as key factors in preventable pigtail catheter malfunctions.
Area of Science:
- Medical Devices
- Emergency Medicine
- Patient Safety
Background:
- Small-caliber chest tubes are frequently used for pneumothorax and pleural effusions.
- Despite being considered less invasive, these tubes exhibit a high complication rate.
- Emergency physicians require a systematic approach to manage complications and device malfunctions.
Purpose of the Study:
- To identify and analyze complications associated with small-caliber chest tubes, specifically pigtail catheters.
- To investigate contributing factors to device malfunction using root cause analysis (RCA).
- To highlight opportunities for improving patient safety in percutaneous catheter procedures.
Main Methods:
- Literature review of small-caliber chest tube complications.
- Analysis of U.S. Food and Drug Administration (FDA) reports for a specific pigtail catheter kit.
- Structured root cause analysis (RCA) of two cases involving retained obturators.
Main Results:
- Common pigtail catheter complications were identified from medical literature.
- Twenty-eight FDA reports detailed kit complications; retained obturator errors were unique to the analyzed cases.
- RCA revealed contributing factors including radiographic clues, human error, communication issues, and device design.
Conclusions:
- RCA identified preventable errors in pigtail catheter insertion, emphasizing the need for improved patient safety.
- Enhanced provider education, systematic procedural approaches, and device design improvements can mitigate risks.
- An end-of-procedure time-out and systematic device assessment are crucial patient safety measures.
More Related Videos
08:51Author Spotlight: Hickman Catheter Use for Long-Term Vascular Access in a Preclinical Swine Model
Published on: March 31, 2023
06:38Use of a Piglet Model for the Study of Anesthetic-induced Developmental Neurotoxicity AIDN: A Translational Neuroscience Approach
Published on: June 11, 2017
Related Concept Videos
Endocarditis IV: Nursing Management
Parentral Nutrition: Centeral and Peripheral Parental Nutrition
PN can be administered through two primary routes:
1. Central Parenteral Nutrition (CPN):
CPN involves delivering a high concentration of nutrients through a large vein. This is typically achieved using a Peripherally Inserted Central Catheter (PICC) or,...
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization IV: Nursing Management