Adam J Graham1, Michele Orini1,2, Ernesto Zacur3
1Barts Heart Centre, Barts Health NHS Trust, London, United Kingdom (A.J.G., M.O., G.D., H.D., N.T.S., C.M., J.L., J.S.M., A.C., J.G., F.P., O.S., S.A., M.L., M.F., M.J.E., A.C., S.S., M.D., R.J.H., R.J.S., P.D.L.).
Background:
ECG imaging (ECGI) has been used to guide treatment of ventricular ectopy and arrhythmias. However, the accuracy of ECGI in localizing the origin of arrhythmias during catheter ablation of ventricular tachycardia (VT) in structurally abnormal hearts remains to be fully validated.
Methods:
During catheter ablation of VT, simultaneous mapping was performed using electroanatomical mapping (CARTO, Biosense-Webster) and ECGI (CardioInsight, Medtronic) in 18 patients. Sites of entrainment, pace-mapping, and termination during ablation were used to define the VT site of origin (SoO). Distance between SoO and the site of earliest activation on ECGI were measured using co-registered geometries from both systems. The accuracy of ECGI versus a 12-lead surface ECG algorithm was compared.
Results:
A total of 29 VTs were available for comparison. Distance between SoO and sites of earliest activation in ECGI was 22.6, 13.9 to 36.2 mm (median, first to third quartile). ECGI mapped VT sites of origin onto the correct AHA segment with higher accuracy than a validated 12-lead ECG algorithm (83.3% versus 38.9%; P=0.015).
Conclusions:
This simultaneous assessment demonstrates that CardioInsight localizes VT circuits with sufficient accuracy to provide a region of interest for targeting mapping for ablation. Resolution is not sufficient to guide discrete radiofrequency lesion delivery via catheter ablation without concomitant use of an electroanatomical mapping system but may be sufficient for segmental ablation with radiotherapy.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
Related Concept Videos
Dysrhythmias V: Evaluating Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...