Related Experiment Video
Updated: Dec 31, 2025

Proximal Cadaveric Femur Preparation for Fracture Strength Testing and Quantitative CT-based Finite Element Analysis
Published on: March 11, 2017
Minimising tibial fracture after unicompartmental knee replacement: A probabilistic finite element study
Elise C Pegg1, Jonathan Walter2, Darryl D D'Lima3
1Centre for Orthopaedic Biomechanics, Department of Mechanical Engineering, University of Bath, UK.
Background:
Periprosthetic tibial fracture after unicompartmental knee replacement is a challenging post-operative complication. Patients have an increased risk of mortality after fracture, the majority undergo further surgery, and the revision operations are less successful. Inappropriate surgical technique increases the risk of fracture, but it is unclear which technical aspects of the surgery are most problematic and no research has been performed on how surgical factors interact.
Methods:
Firstly, this study quantified the typical variance in surgical cuts made during unicompartmental knee replacement (determined from bones prepared by surgeons during an instructional course). Secondly, these measured distributions were used to create a probabilistic finite element model of the tibia after replacement. A thousand finite element models were created using the Monte Carlo method, representing 1000 virtual operations, and the risk of tibial fracture was assessed.
Findings:
Multivariate linear regression of the results showed that excessive resection depth and making the vertical cut too deep posteriorly increased the risk of fracture. These two parameters also had high variability in the prepared synthetic bones. The regression equation calculated the risk of fracture from three cut parameters (resection depth, vertical and horizonal posterior cuts) and fit the model results with 90% correlation.
Interpretation:
This study introduces for the first time the application of a probabilistic approach to predict the aetiology of fracture after unicompartmental knee replacement, providing unique insight into the relative importance of surgical saw cut variations. Targeted changes to operative technique can now be considered to seek to reduce the risk of periprosthetic fracture.
More Related Videos
08:20A Reliable and Reproducible Critical-Sized Segmental Femoral Defect Model in Rats Stabilized with a Custom External Fixator
Published on: March 24, 2019
07:33In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023