Comparison of cuff-based and cuffless continuous blood pressure measurements in children and adolescents

Jacek Zachwieja1, Anna Neyman-Bartkowiak1, Alina Rabiega1

  • 1Department of Pediatric Nephrology and Hypertension, Poznan University of Medical Sciences , Poznan, Poland.

Insights

This study compared cuffless blood pressure (BP) devices using pulse transit time (PTT) with standard cuff-based measurements in children. While correlated, the cuffless device showed higher BP readings and results are not directly interchangeable with traditional methods.

Area of Science:

  • Biomedical Engineering
  • Cardiovascular Research
  • Pediatric Health

Background:

  • Emerging cuffless blood pressure (BP) measurement devices utilize pulse transit time (PTT).
  • The clinical comparability of these novel devices with standard ambulatory BP monitoring in pediatric populations remains largely uninvestigated.

Purpose of the Study:

  • To evaluate the agreement between BP measurements obtained from a cuffless PTT device and a standard cuff-based ambulatory BP monitoring (ABPM) device in children.
  • To determine if BP values from the cuffless device are interchangeable with traditional cuff-based measurements.

Main Methods:

  • Prospective study involving thirty children.
  • Simultaneous BP measurements were taken using a cuffless PTT device (Somnotouch-NIBP) and a standard cuff-based device (Omron 907) on both arms.

Main Results:

  • Significant positive correlations were observed between the cuffless PTT device and standard cuff-based measurements for both systolic BP (SBP) and diastolic BP (DBP).
  • The cuffless device yielded higher mean SBP (127.48 ± 15.98 mmHg) and DBP (68.52 ± 12.36 mmHg) compared to the standard device (SBP: 123.47 ± 14.91 mmHg; DBP: 66.88 ± 11.86 mmHg).
  • Differences in SBP between the methods could reach up to 20 mmHg.

Conclusions:

  • The developed pulse wave velocity-BP (PWV-BP) function demonstrates a significant correlation with sphygmomanometry-derived SBP.
  • Mean 24-hour, awake, and asleep BP values from the cuffless device are not directly interchangeable with cuff-based ABPM values.
  • Despite observed differences, cuffless BP monitoring systems offer promising novel solutions for various clinical applications.
Abstract

Related Concept Videos

Sites for measruring blood pressure01:21

Sites for measruring blood pressure

Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
3.2K
Special considerations while measuring blood pressure01:28

Special considerations while measuring blood pressure

When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
1.1K
Pre-Procedural Guidelines for Assessing Blood Pressure01:10

Pre-Procedural Guidelines for Assessing Blood Pressure

Accurate blood pressure assessment is crucial for diagnosing and managing various health conditions. To ensure the reliability of these measurements, healthcare professionals must adhere to standardized pre-procedural guidelines. These guidelines enhance patient safety and improve the overall quality of healthcare. The following steps are essential for obtaining accurate and consistent blood pressure readings, from using the appropriate tools to ensuring effective communication with the...
772
Assessment of blood pressure in brachial artery(two-step method)01:23

Assessment of blood pressure in brachial artery(two-step method)

Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
1.4K
Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
2.9K
Measurement of Blood Pressure01:17

Measurement of Blood Pressure

Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
2.5K