Effective Continuous Positive Airway Pressure Changes Related to Sleep Stage and Body Position in Obstructive Sleep
Eun Mi Lee1, Tae Hoon Lee2, Ol Lim Park1
1Department of Neurology, Ulsan University Hospital, University of Ulsan College of Medicine, Ulsan, Korea.
Background And Purpose:
The aim of this study was to determine how the sleep stage and body position influence the effective pressure (Peff) in standard upward titration and experimental downward titration.
Methods:
This study applied successful manual titration of continuous positive airway pressure over 3 hours [including at least 15 min in supine rapid eye movement (REM) sleep] followed by consecutive downward titration for at least 1 hour to 22 patients with moderate-to-severe obstructive sleep apnea. We analyzed baseline polysomnography variables and compared the effective pressures (Peff1upward and Peff2downward) between non-REM and REM sleep and between supine and lateral positions using the paired t-test or Wilcoxon signed-rank test.
Results:
During upward titration, Peff1 increased during REM sleep compared to non-REM sleep [9.5±2.9 vs. 8.9±2.7 cm H₂O (mean±SD), ΔPeff1REM-non-REM=0.6±1.1 cm H₂O; p=0.024]. During downward titration, Peff2 was higher in a supine than a lateral position (7.3±1.7 vs. 4.8±1.5 cm H₂O, ΔPeff2supine-lateral=2.5±1.3 cm H₂O; p=0.068). When comparing both upward and downward titration conditions, we found that Peff2 was significantly lower than Peff1 in all sleep stages, especially during REM sleep (Peff1REM vs. Peff2REM=9.5±2.9 vs. 7.4±3.3 cm H₂O) with an overall difference of 2.1±1.7 cm H₂O (p<0.001). Peff in supine sleep decreased from 9.4±3.0 cm H₂O (Peff1supine) to 7.6±3.3 cm H₂O (Peff2supine), with an overall difference of 1.8±1.6 cm H₂O (p<0.001).
Conclusions:
This study has revealed that the collapsibility of the upper airway is influenced by sleep stage and body position. After achieving an initial Peff1, a lower pressure was acceptable to maintain airway patency during the rest of the sleep. The observed pressure decrease may support the use of an automated titration device that integrates real-time positional and sleep-stage factors, and the use of a lower pressure may improve fixed-pressure-related intolerance.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Sleep Apnea
The condition is more prevalent among...
Mechanical Ventilation I: Indication and Settings
Cardiopulmonary Resuscitation II: ACLS Airway Management
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Hyperpnea and Hyperventilation