Related Experiment Video
Updated: Dec 30, 2025

Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
PSEUDO-3D IMRT VERIFICATION WITH EBT3 RADIOCHROMIC FILM
Tereza Hanušová1, Ivana Horáková2, Irena Koniarová1,2
1Department of Dosimetry and Application of Ionizing Radiation, Faculty of Nuclear Sciences and Physical Engineering, Czech Technical University in Prague, Břehová 7, 115 19 Praha 1, Czech Republic.
Abstract:
This work proposes a new method for pseudo-3D verification of intensity-modulated radiation therapy (IMRT) dose distributions. Unlike commercial solutions, it uses measured doses only for gamma evaluation. Its resolution is far better than with electronic detectors within the measured plane and comparable in other directions. It is readily available at clinics because it uses existing resources-a slab phantom and EBT3 films. The method was tested on six IMRT clinical cases. An in-house code for 2D and pseudo-3D gamma analysis was written in MATLAB and compared to OmniPro I'mRT.

