Microbiological surveillance of endoscopes and implications for current reprocessing procedures adopted by an Italian
A Cottarelli1, M De Giusti2, A G Solimini1
1Department of Public Health and Infectious Diseases, Sapienza University of Rome, Italy.
Background:
Hospital acquired infections have been associated with the contamination of flexible endoscopes caused by a failure of the reprocessing procedure. Microbiological surveillance of endoscope reprocessing is valuable for assessing contamination by pathogens. The aim of this study is to evaluate microbiological contamination of endoscopes after reprocessing, and the involvement of reprocessing procedures adopted in endoscopy units of an Italian teaching-hospital.
Methods:
The study was carried out, on several dates in 2014, in 11 endoscopic operation units equipped with 100 endoscopes (18 bronchoscopes, 41 gastroduodenoscopes, 29 colonoscopes, 12 laryngoscopes) and 9 Automated Endoscope Reprocessors. Presence/absence of common pathogens and indicator micro-organisms (including multi-drug resistant bacteria) and Total Microbiological Count (TMC) were obtained from the biopsy channels of endoscopes after reprocessing, from final rinse water of automated endoscope reprocessors and from tap water applying standard microbiological culture methods. Following the European Guidelines for quality assurance in reprocessing, the post-reprocessing criteria were "absence of indicator micro-organisms and absence of TMC in samples obtained from endoscopes' channels".
Results:
A total of 180 samples were collected (143 endoscopes, 25 Automated Endoscope Reprocessors and 12 water supply). Compliance to the European Guidelines was achieved for 112 out of the 180 (62.2%) samples analyzed. Presence of indicator micro-organisms (mainly Klebsiella pneumoniae, Escherichia coli, Pseudomonas aeruginosa and other Gram-negative non-fermenting bacteria) was found in 51 out of 143 endoscopes (35.7%). Multi-drug resistant bacteria were also found. Presence of pathogen micro-organisms was statistically associated with the increase of TMC level, but not with time after reprocessing.
Conclusion:
The study provides information about the microbiological quality of endoscope reprocessing procedures adopted by different endoscopic operation units. The high prevalence of contaminated endoscopes provides evidence of the need to improve the quality of reprocessing.
Related Concept Videos
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Endoscopic Procedures III: Video Capsule Endoscopy
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...