Home-Based Monitor for Gait and Activity Analysis
Published on: August 8, 2019
Cuff-Less Blood Pressure Monitoring with a 3-Axis Accelerometer
Objective:
The aim was to propose a cuff-less, cost-efficient, and ultra-convenient blood pressure monitoring technique with a 3-axis accelerometer.
Methods:
The efficacy of the proposed approach was examined in 8 young healthy volunteers undergoing different activities with a 3-axis accelerometer leveled on their upper chest. The 3-dimensional accelerations were exploited to select features for the calculation of systolic pressure (SP) and diastolic pressure (DP); the whole process involved signal processing, feature extraction, linear multivariate regression, and leave-one-out cross validations (LOOCV).
Results:
DP and SP could be approximated with the linear combination of the extracted features: the L2 norm of lateral acceleration for both DP and SP, state variation (defined in the proposed algorithm) of vertical acceleration for SP, and I-J interval (defined in ballistocardiogram) of vertical acceleration for DP. The correlation coefficient (r) of the estimated and the measured DP was 0.97, and for SP, r = 0.96. In LOOCV, our best validated results in difference errors were -0.02±3.82 mmHg for DP and -0.59 ± 7.46 mmHg for SP.
Conclusion:
Compared to AAMI criteria, the proposed acceleration-based technique fulfilled the requirement. The accelerometer-based technique showed the potential to monitor blood pressure cuff-lessly, cost-efficiently, ultra-conveniently, and to be embedded in a long-term wearable device for clinical usage.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: