Manufacturing Abdominal Aorta Hydrogel Tissue-Mimicking Phantoms for Ultrasound Elastography Validation
Published on: September 19, 2018
Quantitative quasi-static ultrasound elastography using reference layer: Ex-vivo study
Abstract:
It is well-documented in the literature that changes in tissue elasticity are generally correlated with disease condition. In the case of diffuse liver disease, the elasticity of the liver reduces progressively. However, this change does not clearly manifest in conventional ultrasound examinations. Although quasi-static elastography is popular in clinical applications where qualitative assessment of relative tissue stiffness is enough, its potential is relatively underutilized in liver imaging due to the need for quantitative stiffness value. Recently, it was demonstrated that using a reference layer of known stiffness, one could produce quantitative modulus elastograms of the target tissue using quasi-static elastography using simulations and phantom experiments. Here, we examined the performance of this approach on ex-vivo goat liver samples and compare the estimated modulus values to that obtained from indentation measurements. The results suggest that using this approach of reference layer yields Young's modulus values within 10% error compared to the ground truth.