Diagnostic performance of modern imaging instruments in glaucoma screening
Elina Karvonen1,2, Katri Stoor3,2, Marja Luodonpää3,2
1Department of Ophthalmology, Oulu University Hospital, Oulu, Finland elina.karvonen@fimnet.fi.
The British Journal of Ophthalmology
|January 18, 2020
Summary
Spectral-domain optical coherence tomography (SD-OCT), scanning laser polarimetry (GDx), and scanning laser ophthalmoscopy (HRT) showed moderate accuracy for glaucoma screening in middle-aged adults. These imaging devices alone are not reliable for screening this population.
Area of Science:
- Ophthalmology
- Medical Imaging
- Public Health
Background:
- Glaucoma screening in middle-aged populations is crucial for early detection and prevention of vision loss.
- Modern imaging devices offer potential for objective glaucoma assessment.
- Evaluating the efficacy of these technologies in unselected populations is essential.
Purpose of the Study:
- To assess the applicability of spectral-domain optical coherence tomography (SD-OCT), scanning laser polarimetry (GDx), and scanning laser ophthalmoscopy (HRT) for glaucoma screening.
- To determine the diagnostic performance of these imaging modalities in a middle-aged, unselected population.
Main Methods:
- A population-based study (Northern Finland Birth Cohort Eye Study) included participants aged 45-49 years.
- Comprehensive eye examinations included SD-OCT (retinal nerve fibre layer and ganglion cell layer analysis), GDx, and HRT.
- Automated classification performance was evaluated against a clinical glaucoma diagnosis (2 out of 3 rule).
Main Results:
- Glaucomatous damage was identified in 1.1% of subjects (43 eyes).
- Sensitivities ranged from 31% (HRT) to 61% (OCT combination), with specificities from 88% (GDx) to 96% (HRT).
- Area under the curve values were moderate (0.73-0.76), and post-test probabilities of glaucoma after a positive finding were low (5-11%).
Conclusions:
- The screening capabilities of OCT, GDx, and HRT were comparable but only moderate in accuracy.
- These imaging parameters, when used alone, are not sufficiently reliable for glaucoma screening in an unselected middle-aged population.
- Further research may be needed to improve the diagnostic performance of these tools for population-based screening.
Related Concept Videos
Glaucoma: Overview
1.2K
Glaucoma is an eye condition characterized by increased intraocular pressure that damages the retina and optic nerve, leading to irreversible blindness if left untreated. The human eye has various components, including the cornea, iris, pupil, lens, and optic nerve. Aqueous humor is secreted by the epithelium of the ciliary body in the posterior chamber and flows through the trabecular meshwork and canal of Schlemm, maintaining normal intraocular pressure. The trabecular meshwork and the canal...
1.2K
Open Angle Glaucoma: Treatment
905
In open-angle glaucoma, the iridocorneal angle remains open, but the trabecular meshwork becomes stiff, slowing down the outflow of aqueous humor. This causes a buildup of aqueous humor in the anterior chamber, leading to a sudden increase in intraocular pressure. The treatment for open-angle glaucoma focuses on reducing the elevated intraocular pressure by either decreasing the secretion of aqueous humor or increasing its outflow.
Drugs such as carbonic anhydrase inhibitors, α2- and...
Drugs such as carbonic anhydrase inhibitors, α2- and...
905