The OralChromaTM CHM-2: a comparison with the OralChromaTM CHM-1
I Laleman1, C Dekeyser2, A Wylleman2
1Section of Periodontology, Department of Oral Health Sciences, KU Leuven & Dentistry, University Hospitals Leuven, Kapucijnenvoer 33, 3000, Leuven, Belgium. isabelle.laleman@kuleuven.be.
Objectives:
Already decades ago, the diagnosis of halitosis was facilitated with the arrival of chair-side instruments to score the breath odor. These devices are used for a more objective assessment of halitosis compared with organoleptic scoring, but these too have their disadvantages. To overcome some of the drawbacks of the original model of the OralChromaTM (CHM-1), few years ago a second generation of this machine (CHM-2) was introduced. This study compared both devices in a clinical setting.
Materials And Methods:
All records of the patients visiting a specialized halitosis consultation over a period of 5 years (2012-2017) were examined. The correlations of the OralChromaTM CHM-1 and CHM-2 with the organoleptic and Halimeter® measurements were analyzed. Additionally the sensitivity, specificity, and negative and positive predicted values were calculated.
Results:
A total of 581 data points were included (CHM-1: 292, CHM-2: 289). The correlations between both models with the organoleptic measurements were not statistically significant different. The CHM-2 seemed superior to the CHM-1 in the quantification of dimethyl sulfide with a detection rate of 95% and 61%, respectively. Additionally, the CHM-2 was significantly more sensitive for dimethyl sulfide than the CHM-1. However the latter showed in turn a better sensitivity for hydrogen sulfide and methyl mercaptan.
Conclusion:
The CHM-2 showed a better sensitivity for dimethyl sulfide than its predecessor. However, its sensitivity for hydrogen sulfide and methyl mercaptan was worse.
Clinical Relevance:
Dimethyl sulfide is the main volatile implicated in extra-oral blood-borne halitosis, this makes the OralChromaTM CHM-2 the instrument of choice when this is suspected.
Related Concept Videos
¹H NMR Chemical Shift Equivalence: Enantiotopic and Diastereotopic Protons
In chiral compounds such as 2-butanol, replacing the methylene hydrogens at C3 produces a pair of...
¹H NMR Chemical Shift Equivalence: Homotopic and Heterotopic Protons
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Chromatographic Methods: Terminology