Introduction:
Reverse osmosis (RO), a major advance in hemodialysis (HD) safety, effectively clears most water organisms. Delftia acidovorans is an environmental water-borne pathogen that is rarely reported to cause human infections. We report a pseudo outbreak caused by colonization of RO machines with D acidovorans with no reported human infections and interventions to improve HD safety.
Methods:
Repeated positive RO product water cultures triggered our hospital to initiate an investigation, RO machines were examined for mechanical integrity by biomedical engineers. Cultures of product water as well as RO parts were done. Testing for bacterial relatedness after identification was performed. An investigation was conducted in a systematic fashion to determine the cause and the extent of the problem.
Results:
Upon formal review of policies and procedures, there were minor deficiencies. Rectifying these deviations from policies did not stop the repeated positive water cultures. A 7-step investigation and correction was successfully conducted. City water testing with filtration and concentration methods was positive for D acidovorans. Major renovation of the HD unit with replacement of all RO machines and a 2-step water filtration resulted in elimination of the pseudo outbreak.
Conclusion:
City water was the source of biofilm formation of D acidovorans in RO machines that was not possible to eradicate. Application of incoming water filters was an effective preventive strategy. Replacement of RO machines after 4 failed disinfection attempts is the most cost-effective strategy as well for persistent positive water cultures. The HD remains RO water culture negative 2 years after changes.
More Related Videos
12:11A Modified EPA Method 1623 that Uses Tangential Flow Hollow-fiber Ultrafiltration and Heat Dissociation Steps to Detect Waterborne Cryptosporidium and Giardia spp.
Published on: July 9, 2012
06:27A Retrograde Implantation Approach for Peritoneal Dialysis Catheter Placement in Mice
Published on: July 20, 2022
Related Concept Videos
Hemodialysis II: Procedure and Complications
Extracorporeal Removal of Drugs: Peritoneal Dialysis and Hemodialysis
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Hemodialysis I: Introduction
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration