Related Experiment Video
Updated: Dec 30, 2025

Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
[Evaluating Eye Lens Dose of Neurovascular and Cardiac Electrophysiology Interventional Physician]
Mamoru Kato1,2, Koichi Chida2, Takato Ishida1
1Department of Radiology and Nuclear Medicine, Akita Cerebrospinal and Cardiovascular Center.
Purpose:
The International Commission on Radiological Protection recommended that interventional radiologies (IRs) have high radiation doses and that staff may also be exposed to high doses. In the present study, we measured the radiation exposure dose [3 mm dose equivalent, Hp (3) ] in the eye using an appropriate dosimeter placed next to the physician' s eye during neurovascular intervention procedure (Neuro-IR) and interventional cardiac electrophysiology procedure (EP-IR).
Method:
Physicians wore a direct eye dosemeter just lateral to the left eye and an additional direct eye dosemeter outside the radiation protective glasses close to their left eye. Additionally, a neck badge [0.07 mm dose equivalent, Hp (0.07) ] was worn outside the protective apron to the left of the neck, to compare the direct eye dosimeter estimated doses. The occupational eye lens dose was evaluated over a period of 6-month.
Results:
The maximum Hp (3) of the Neuro-IR physician was estimated 5.1 mSv without the radiation protective glasses and 1.6 mSv with the radiation protective glasses. On the other hand, the maximum Hp (3) of the EP-IR physician was estimated 29 mSv without the radiation protective glasses and 15 mSv with the radiation protective glasses.
Conclusion:
Physicians eye lens dose [Hp (3) ] tended to be overestimated by the neck badge measurements [Hp (0.07)]. A correct evaluation of the lens dose [Hp (3) ] using the direct eye dosimeter is recommended. Although we found a positive correlation between Hp (0.07) and Hp (3), the value of R2 in the regression equation is low, we recommended that the eye lens dose estimated carefully from Hp (0.07).
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
11:27Magnetically-Assisted Remote Controlled Microcatheter Tip Deflection under Magnetic Resonance Imaging
Published on: April 4, 2013
Related Concept Videos
Cardiac Catheterization IV: Nursing Management
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization I: Pre-Procedure Overview