Measurement of skin hydration with a portable device (SkinUp® Beauty Device) and comparison with the Corneometer®
Thais V A Westermann1, Vinicius Rodrigues Viana2, Clemilson Berto Junior3
1College of Pharmacy, Federal University of Rio de Janeiro, Macaé, Brazil.
Background:
Epidermis hydration measurement is an important tool to assess its health. SkinUp® is a portable device that measures moisture and oil levels of skin through impedance method, it is cheap and has small dimensions, but this equipment has not yet been cited in the scientific literature. Thus, the objective of this work was to validate SkinUp® equipment by comparing its results with Corneometer® CM825.
Materials And Methods:
Twenty subjects had skin hydration measurement on forearm, cheeks, and forehead. Measurements obtained with the two instruments were correlated by calculating the Pearson correlation coefficient and the P-value significance. The sensitivity of the equipment to different regions and treatments and their accuracy were also evaluated.
Results:
Positive correlations between equipment were found for skin hydration, with P ≤ .001. Both instruments presented high sensitivity to the different treatments (P < .001). When sensitivity to different regions was evaluated, both showed greater hydration in the cheek and forehead when compared to the forearm (P < .005). The measurements between the cheek and forehead showed no significant difference when compared to each other. In addition, the two devices have good repeatability in the measurements.
Conclusion:
These results indicate that SkinUp® can be used in the future for skin analysis in in vivo assays.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: