Time series evaluation of improvement interventions to reduce alarm notifications in a paediatric hospital

Colleen M Pater1, Tina K Sosa2, Jacquelyn Boyer3

  • 1The Heart Institute, Cincinnati Children's Hospital Medical Center, Cincinnati, Ohio, USA colleen.pater@cchmc.org.

BMJ Quality & Safety
|January 22, 2020
PubMed

Insights

This study reduced non-actionable alarm notifications in a pediatric cardiology unit by 68% using quality improvement methods. This enhanced patient safety and significantly improved nursing satisfaction with alarm response.

Area of Science:

  • Healthcare Quality Improvement
  • Patient Safety
  • Clinical Workflow Optimization

Background:

  • The Joint Commission highlighted inpatient alarm reduction as crucial for patient safety, satisfaction, and workflow.
  • Excessive non-actionable alarms disrupt care and decrease provider responsiveness.

Purpose of the Study:

  • To decrease alarm notifications per monitored bed per day by 60% in a pediatric tertiary care center's acute care cardiology unit.
  • To improve nursing satisfaction with alarm response.
  • To maintain patient safety throughout the intervention.

Main Methods:

  • A multidisciplinary team implemented quality improvement (QI) methods, including updating technology, refining alarm logic, and modifying bedside workflow.
  • Key interventions involved silencing bedside monitor volumes and establishing new alarm protocols.
  • Plan-Do-Study-Act (PDSA) cycles guided the iterative improvement process.

Main Results:

  • Initial alarm notifications per monitored bed per day decreased by 68%, from 71 to 22, over 3.5 years.
  • Nursing satisfaction with alarm response significantly improved, with agreement rising from 32% to 76%.
  • Improvements were sustained for over a year without adverse effects on patient safety measures.

Conclusions:

  • Quality improvement methods and technological advancements can effectively reduce alarm notifications in pediatric settings.
  • Sustained reduction in alarms enhances patient safety and nursing satisfaction.