Hemolysis. A comparative study of four nonpulsatile pumps
1Cleveland Clinic Foundation, Ohio.
Summary
This study compared hemolysis in four nonpulsatile blood pumps. The Biopump BP-80 showed the lowest hemolysis index, indicating superior performance in blood handling.
Area of Science:
- Biomedical Engineering
- Cardiovascular Devices
- Hemodynamics
Background:
- Hemolysis is a critical performance metric for blood pumps.
- Inconsistent testing methodologies hinder direct comparison of published hemolysis data.
Purpose of the Study:
- To evaluate and compare the hemolysis generated by four nonpulsatile blood pumps under identical conditions.
- To establish a standardized method for assessing blood pump performance regarding hemolysis.
Main Methods:
- Four nonpulsatile blood pumps (Roller, Biopump Model 600, Biopump BP-80, Hemadyne) were tested in a standardized closed circuit.
- Tests were conducted using calf blood at 37°C, pumped at 6 L/min for 12 hours against a 100 mmHg afterload.
- The hemolysis index (HI) was calculated based on hemoglobin release per 100 L pumped.
Main Results:
- The mean hemolysis index (HI) varied significantly among the tested pumps.
- The roller pump exhibited the highest HI (0.0467), while the Biopump BP-80 showed the lowest HI (0.0076).
- No statistically significant difference in HI was observed among the three centrifugal pumps (Biopump Model 600, Biopump BP-80, Hemadyne).
Conclusions:
- Standardized testing reveals significant differences in hemolysis among nonpulsatile blood pumps.
- The Biopump BP-80 demonstrated the most favorable blood compatibility among the tested devices.
- Centrifugal blood pumps, in general, may offer comparable hemocompatibility when tested under consistent conditions.