A novel disposable sensor for measure intra-abdominal pressure
José S Camacho-Juárez1, Bersaín Alexander-Reyes1, Antonio Morante-Lezama1
1Facultad de Ciencias, Universidad Autónoma de San Luis Potosí, San Luis Potosí. México.
A novel disposable sensor for intra-abdominal pressure (IAP) measurement, the DSIAP, shows high accuracy in vitro and clinical settings. This new method overcomes limitations of the traditional Kron
Area of Science:
- Medical Devices
- Physiological Monitoring
- Critical Care Medicine
Background:
- Intra-abdominal pressure (IAP) measurement is crucial in patient care.
- The conventional Kron's method for IAP measurement has drawbacks, including patient discomfort due to bladder water infusion.
Purpose of the Study:
- To introduce and validate a novel disposable sensor for intra-abdominal pressure (DSIAP) measurement.
- To address the limitations associated with the Kron's method.
Main Methods:
- The DSIAP sensor was evaluated through in vitro and clinical testing.
- Comparison with Kron's method utilized Pearson correlation and Bland-Altman analysis.
- In vitro testing involved 159 simulated IAP measurements; clinical testing included 20 pairs of measurements from ICU patients.
Main Results:
- In vitro tests demonstrated a strong correlation (r = 0.99) between DSIAP and reference measurements.
- Clinical tests showed a very good correlation (r = 0.973) between DSIAP and Kron's method.
- Both in vitro and clinical results indicated minimal bias and narrow confidence intervals, suggesting high accuracy and reliability.
Conclusions:
- The DSIAP sensor exhibits performance comparable to existing clinical pressure transducers.
- The DSIAP effectively overcomes the limitations of the Kron's method for IAP monitoring.
- This novel disposable sensor offers a promising alternative for accurate and less invasive IAP measurement.